Skip to main content

Table 1 Accuracy and precision of linear detection range of PCR platforms

From: Applicability of RNA standards for evaluating RT-qPCR assays and platforms

 

Slope

R2

Platform

Dynamic arrays

ABI 7900HT

Dynamic arrays

ABI 7900HT

Template

Pre-amplified

cDNA

cDNA

cDNA

Pre-amplified

cDNA

cDNA

cDNA

ERCC-

      

13

-1.06

-1.08

-1.06

0.999

0.993

0.997

42

-1.07

-1.11

-1.06

0.998

0.970

0.995

81

-1.02

-1.07

-1.04

0.998

0.976

0.998

84

-1.03

-1.07

-1.01

0.998

0.973

0.996

95

-1.07

-1.12

-1.08

0.997

0.975

0.998

99

-1.06

-1.06

-1.05

0.998

0.985

0.997

113

-1.06

-1.15

-1.04

0.998

0.971

0.997

171

-1.12

-1.13

-1.08

0.998

0.982

0.996

All

-1.06

-1.10

-1.05

0.999

0.978

0.997

  1. Linear regression was performed with Ct values vs. log2(ERCC RNA copy number) for each PCR platform across the linear range marked in Figure 1. The mean slope and R2 values from three independent qPCR experiments reflect the accuracy and precision of the detection of the 10-fold differences in copy numbers between samples.