Skip to main content

Table 4 Analytical accuracy

From: Analytical validation of the Oncotype DX prostate cancer assay – a clinical RT-PCR assay optimized for prostate needle biopsies

Gene Assay

Average % Bias

ARF1

3.0

ATP5E

9.7

AZGP1

1.8

BGN

3.6

CLTC

−0.3

COL1A1

1.1

FAM13C1

−0.8

FLNC

1.2

GPS1

4.9

GSN

4.8

GSTM2

3.1

KLK2

2.2

PGK1

1.9

SFRP4

3.9

SRD5A2

−1.9

TPM2

1.0

TPX2

−2.0

  1. The estimated analytical accuracy of predicted RNA concentrations relative to the known input RNA concentrations for each of the 17 Oncotype DX Prostate Cancer Assay gene assays; the average % bias is taken over the nominal, ½ and ¼ qPCR input levels.