Table 4 Input DNA copy number and genotype call accuracy

From: High-throughput single nucleotide polymorphism genotyping using nanofluidic Dynamic Arrays

    Copy number per chamber
    0.9 4.5 9 45 90
A1A1 Genotype Count No Call 50 2    
   YY 73 121 123 123 123
  Call rate   59.3% 98.4% 100% 100% 100%
  % Error   0% 0% 0% 0% 0%
A1A2 Genotype Count No Call 72 4 7   
   XX 15 18 16   
   XY 8 80 89 123 123
   YY 28 21 11   
  Call rate   41.5% 96.7% 94.3% 100% 100%
  % Error   35% 32% 22% 0% 0%
A2A2 Genotype Count No Call 39 4    
   XX 84 119 123 123 123
  Call rate   68.3% 96.7% 100% 100% 100%
  % Error   0% 0% 0% 0% 0%
Grand No Call    161 10 7 0 0
Grand Error    43 39 27 0 0
Grand Call Rate    56.4% 97.3% 98.1% 100% 100%
Grand % Error    11.7% 10.6% 7.3% 0.0% 0.0%