Comparison of the Illumina NextSeq 2000 and GeneMind Genolab M sequencing platforms for spatial transcriptomics

Abstract

Background

The Illumina sequencing systems demonstrate high efficiency and power and remain the most popular platforms. Platforms with similar throughput and quality profiles but lower costs are under intensive development. In this study, we compared two platforms Illumina NextSeq 2000 and GeneMind Genolab M for 10x Genomics Visium spatial transcriptomics.

Results

The performed comparison demonstrates that GeneMind Genolab M sequencing platform produces highly consistent with Illumina NextSeq 2000 sequencing results. Both platforms have similar performance in terms of sequencing quality and detection of UMI, spatial barcode, and probe sequence. Raw read mapping and following read counting produced highly comparable results that is confirmed by quality control metrics and strong correlation between expression profiles in the same tissue spots. Downstream analysis including dimension reduction and clustering demonstrated similar results, and differential gene expression analysis predominantly detected the same genes for both platforms.

Conclusions

GeneMind Genolab M instrument is similar to Illumina sequencing efficacy and is suitable for 10x Genomics Visium spatial transcriptomics.

Peer Review reports

Background

Single-cell RNA sequencing (scRNA-seq) has become a powerful approach to characterize the gene expression profile in single cells [1]. This method allows processing tens and hundreds of thousands of single cells simultaneously to measure their transcriptional profiles [2, 3]. ScRNA-seq technique provides a higher resolution of cellular differences and better understanding of cell populations and their relationships [4, 5]. However, a serious practical obstacles related to cell stress, cell death, and/or cell aggregation during cell isolation for scRNA-seq exist, accompanied by the loss of spatial context. Moreover, some cell types in tissue, in particular immune cells, are not always easily isolated from tissue [6].

Spatial transcriptomics provides quantitative gene expression data within the spatial context of tissues and cells on tissue sections [7]. Among single-cell spatial transcriptomics methods, 10x Genomics Visium is a barcoding technique that gives an information about cell-type composition on the spots [8]. While “bulk” cDNA libraries have relatively plain structure, Visium cDNA libraries are composed of several elements making these libraries more sophisticated. For example, Visium libraries obtained from formalin-fixed paraffin-embedded (FFPE) samples include spatial barcodes, unique molecular identifiers (UMI) at the Read 1, and probe sequences at the Read 2. Base calling errors at these library elements could cause loss of information about spatial position, UMI assignment, and probe origin. Therefore, these complicated libraries demand high sequencing quality.

The Illumina sequencing systems demonstrate high efficiency and power and remain the most popular platforms that is confirmed by the dominance in the literature and in the amount of data submitted to the Sequence Read Archive [9]. Platforms with similar throughput and quality profiles but lower costs (cost per read) are under intensive development [10].

In October 2020, GeneMind Biosciences Company Limited (GeneMind) launched a new sequencing instrument (GenoLab M) that is fully compatible with Illumina based libraries and does not require any library conversion [11]. It was revealed that the GenoLab M is a promising sequencing platform for transcriptomics and long non-coding RNA profiling in animal, plant, and human with comparable performance but a lower cost compared to NovaSeq 6000 (Illumina) [11, 12]. However, the performance of the GenoLab M platform in other applications has not yet been tested, for example, for single cell sequencing including spatial transcriptomics.

Here, we compared the relative performance of GeneMind Genolab M and Illumina NextSeq 2000 for 10x Genomics Visium spatial sequencing. The same cDNA libraries prepared from three ovarian cancer samples were sequenced in two platforms, and the efficacy and accuracy of results were analyzed by bioinformatics.

Results

Quality of sequencing

In order to assess concordance of the Illumina NextSeq 2000 and GeneMind Genolab M sequencing platforms for spatial transcriptomics, we sequenced three 10x Genomics Visium libraries of FFPE ovarian tumor tissues. The main metrics are provided in Table 1.

Table 1 Unity of sequencing quality of the samples in two sequencing platforms

The mean number of reads for the NextSeq 2000 and Genolab M platforms was about 72 and 62 million, respectively. The percentage of valid barcodes and percentage of valid UMIs were 0.74% and 0.1% higher in NextSeq 2000. Base calling accuracy was also higher in the NextSeq 2000. The percentage of Q30 bases (99.9% base call accuracy) in barcode, probe, and UMI reads was about 1.6%, 2.2%, and 2.6% higher in the NextSeq 2000, respectively. Furthermore, there was apparent difference in the sequencing saturation, which was higher when using NextSeq 2000. This could be due to the fact that the sequencing depth and the number of read duplication were lower with using the Genolab M platform (Supplementary Table 1).

Raw reads mapping and counting

The raw sequencing reads from two sequencers were aligned to the probe reference and assigned to tissue spots by corresponding barcode sequences, and then the aligned and assigned reads were counted. The summarized quality metrics of processed data are included in Table 2. The fraction of reads under tissue spots was consistent between two platforms with insignificant differences. The mean number of reads per spot and the mean number of reads per spot under tissue were in accordance with the sequencing depth. The median number of UMI counts and genes per spot and the number of detected genes were higher in the Genolab M platform, excepting the number of detected genes in B1-2 slide. However, uniquely detected genes in both platforms had markedly low number of UMIs (Supplementary Figure S1A, B) suggesting that these genes have sporadic origin. The percentage of confidently mapped reads to probe set was slightly higher for NextSeq 2000 that is concordant with higher sequencing quality of the probe reads. Nevertheless, Genolab M showed the higher percentage of confidently mapped reads to filtered probe set.

Table 2 Quality control metrics of the raw read mapping and counting

The filtered probe set is the probe set filtered from hybridization probes with high similarity to potential non-specific target genes, so reads from this set are mapped to unique genes. However, the differences between read mapping to the filtered probe set were insubstantial and could arise sporadically.

Characterization of the genes, UMIs, reads, and tissue spots

The raw sequencing reads were converted into the feature-barcode matrix following a read mapping and counting in the Space Ranger. The resulting feature-barcode matrixes were loaded into the R environment for downstream analysis in the Seurat R package. The Pearson correlation coefficient was used to assess concordance of the sequencing results in the tissue spots. There were strong correlations between the total number of UMIs and the total number of detected genes in the tissue spots (Fig. 1A). The gene-UMI relationship (Fig. 1B) for all samples was consistent between two platforms and in accordance to expected distribution. We surveyed GC-content profiles in both forward and reverse reads. GC-content profiles were highly consistent between two platforms (Fig. 1C), indicating no any apparent biases between read coverage in two sequencers.

Fig. 1
figure 1

Characterization of the read counts, UMIs, and detected genes obtained from two sequencing platforms. A Correlation between the number of transcripts (UMIs) and the number of detected genes in the tissue spots. B The genes and UMIs distribution in the tissue spots in two sequencing platforms. C GC-content profiles of raw reads in two sequencing platforms. D The unique and common genes between two sequencing platforms. E Distribution of the Pearson correlation coefficients of the common SCT transformed gene counts

In order to perform downstream analysis, we filtered out low expressed genes and tissue spots with insufficient number of genes using the following thresholds: genes expressed in less than 10 tissue spots and tissue spots with less than 200 filtered genes. Then, overlapping of the filtered genes between two platforms was examined (Fig. 1D). Since there is no well-established criterion to filter out low-expressed genes and tissue spots, different studies choose different thresholds and criterions. The main purpose of filtering is removing low expressed genes, which can affect dimension reduction processes introducing additional noise to data. The thresholds in our manuscript were chosen by survey articles where 10x Genomics Visium technology was used. Closest thresholds to our analysis were used within the article [13].

The number of unique genes in the Genolab M platform was still higher than in the NextSeq 2000, while the proportion between unique and overlapping genes after filtration in both platforms became lower.

Nevertheless, more detailed insight revealed that all unique genes in both platforms had poor coverage and distributed on the edge of the chosen threshold for gene filtration (Supplementary Figure S1C) that indicate their sporadic nature. The read counts were normalized via the SCTransform function in the Seurat package. The SCTransform utilized regularized negative-binomial regression to remove the influence of sequencing depth and other specified unwanted sources of variation. The normalized counts were used for correlation analysis between the overlapping genes (Fig. 1D). There were strong correlations between overlapping genes in spots sequenced on both platforms. Mean correlation coefficients were 0.82, 0.90, and 0.99 in A1-1, A1-2, and B1-2 samples, respectively (Fig. 1E). Analysis of the sequencing quality metrics did not reveal any robust metrics explaining differences in correlation coefficients obtained for the samples.

Dimension reduction, clusterization, and differential expression

The normalized read counts were merged and renormalized via the SCTransform function. Figure 2A demonstrates no observable batch effect between samples sequenced on NextSeq 2000 and Genolab M. In further analysis, SCTransform function without batch effect correction was applied. The resulted SCT normalized counts were used for linear dimension reduction via PCA approach, and the first 30 principal components were used for clusterization and non-linear dimension reduction. SNN analysis based clustering algorithm was used for clusterization of tissue spots. UMAP approach was used for visualization of tissue spots and clusterization in two dimensions. Results of SNN clusterization and UMAP dimension reduction are depicted in Fig. 2A. Tissue spots from the samples analyzed by different sequencing platforms were placed in close proximity. Figure 2B demonstrates cluster distribution on the tissue sections. There were not recognizable differences between spatial cluster distribution in spots sequenced on both platforms. A1-1 sample had the most complex structure regarding number of defined clusters and their distribution on the tissue section. A1-2 sample had less number of clusters, and B1-2 section had simplest spatial distribution of clusters. Collectively, aforementioned evidence regarding spatial clustering could account for difference in tissue spot correlation coefficients between three samples.

Fig. 2
figure 2

Downstream non-linear dimension reduction, clusterization, and differential expression of the filtered and normalized data. A The UMAP dimension reduction and clusterization results of the sequencing counts from two sequencing platforms without batch effect correction. B Spatial distribution of the obtained clusters. C-E Venn diagrams of the common and the platform-unique differentially expressed genes (DEGs) of the slide clusters

Since there is no observable batch effect, DEG analysis can be performed directly on counts after sequencing depth normalization via SCTransform. The SCT normalized counts were used in the Wilcoxon Rank Sum test to find differentially expressed genes in each sample separately. Genes were considered differentially expressed if FDR < 0.01 and LFC > 0.25. Figure 2C-E demonstrates results of differential expression analysis. The number of unique genes detected by NextSeq 2000 and Genolab M is around 16% out of all DEGs detected by both platforms for all clusters; the overlapping rate is about 68%. Therefore, DEGs predominantly overlapped between same clusters originated from two sequencing platforms.

Characterization of differentially expressed genes

The unique and common DEGs were further characterized to evaluate origin of the platform-specific DEGs. The distribution of DEG counts in tissue spot clusters was visualized to evaluate platform-specific DEGs. Supplementary Figure S2 demonstrates that both unique and common genes had corresponding count coverage; therefore, platform-specific DEGs did not arise from low expressed genes. The distributions of LFC and FDR of DEGs were visualized following count coverage characterization of DEGs. Figure 3A demonstrates distribution of DEG LFC in tissue spot clusters. Predominantly, all platform-specific DEGs had LFC distributed on the edge of the chosen threshold. Furthermore, FDR of platform-specific DEGs (Fig. 3B) tended to have least significant values. The LFC and FDR distributions indicated that platform-specific DEGs predominantly had lowest –log (FDR) and LFC close to the chosen LFC threshold. Therefore, unique DEGs could be primarily arose due to stochastically exceeding in chosen thresholds in both platforms. On the contrary, common DEGs had steadily higher -log(FDR) and LFC compared to unique DEGs. Supplementary Figure S3 demonstrates distinct differences in FDR and LFC between common and unique DEGs detected in clusters of sequenced samples. Top DEGs almost fully consist of overlapping DEGs.

Fig. 3
figure 3

Cluster DEG characterization by the log twofold-change (LFC) and the FDR. A LFC distribution of the common and unique DEGs in the clusters. B FDR distribution of the common and unique DEGs in the clusters. Count is a number of DEGs

Discussion

This is the first study demonstrating the use of GeneMind Genolab M platform for 10x Genomics Visium spatial transcriptomics. Moreover, this study first compared sequencing performance of GeneMind Genolab M with Illumina NextSeq 2000 platform. Both platforms employ sequencing by synthesis using a reversible terminator approach [14]. Fluorescently labeled nucleotides are incorporated at a time by polymerase to extending sequencing primer each sequencing cycle. Differences between Genolab M and NextSeq 2000 sequencing techniques include distinct fluorescent labeling approaches. Genolab M utilizes four-color fluorescent labeling marking each nucleotide with distinct fluorescent labels. NextSeq 2000 employs two fluorescent dyes in four combinations: one label for C, another label for T, both labels for A, and no label for G. More specified protocol for GenoLab M sequencing is described in the study by Liu et al. [12].

Our study showed that GeneMind Genolab M sequencing platform generates consistent with Illumina NextSeq 2000 sequencing results. Sequencing quality metrics indicated highly comparable base calling accuracy in UMI sequence, spatial barcode sequence, and probe read sequence. Comparable base calling accuracy resulted in consistent quality control metrics of the raw read mapping and counting. Furthermore, Genolab M and NextSeq 2000 platforms had similar levels of RNA molecule (UMI) and gene detection. Both platforms had identical GC-content profiles of raw reads. Almost all detected genes were overlapped between two platforms; moreover, there were strong correlations between these overlapping genes in spots. Normalization, dimension reduction and clusterization showed particularly consistent results. Identical spot clusters were identified in each sample sequenced using both platforms. Spatial distribution of spot clusters was predominantly similar without obvious differences. The concordance of both platforms was also high in the DEG analysis between obtained clusters. Furthermore, platform-specific DEGs predominantly had lowest FDR and LFC; therefore, unique DEGs could be primarily arose stochastically. Meanwhile, common DEGs detected in clusters of sequenced samples have the highest –log (FDR) and LFC and almost fully consist of top DEGs. Since top DEGs are almost fully overlapped between both platforms, downstream analysis, e.g. gene set enrichment analysis, should produce highly consistent results. All above-mentioned data indicate high concordance between GeneMind Genolab M and Illumina NextSeq 2000 platforms for 10x Genomics Visium spatial sequencing.

Conclusions

In summary, this is the first study demonstrating that GeneMind Genolab M instrument has similar to Illumina NextSeq 2000 sequencing efficacy and is suitable for 10 × Genomics Visium spatial transcriptomics.

Materials and methods

Sample description and library preparation

Three tumor FFPE samples obtained from patients with high-grade serous ovarian cancer were analyzed. Patients were treated in the Department of Gynecological Oncology, Cancer Research Institute of Tomsk National Research Medical Center (Tomsk, Russia). The study was carried out according to Declaration of Helsinki (from 1964, revised in 1975 and 1983) and was approved by the local committee of Medical Ethics of Tomsk Cancer Research Institute; all patients signed informed consent for the study. Shortly, FFPE tissue sections were placed on the Capture Areas of 10x Genomics Visium slides, deparaffinized, and stained by hematoxylin and eosin. Tumor tissue slides were scanned using the Leica Aperio AT2 station (Leica, Germany) and images were processed with Aperio ScanScope. Libraries were prepared using standard protocol provided by 10x Genomics (User Guide | CG000407).

Sequencing

The obtained cDNA libraries were analyzed following 10x Genomics sequencing recommendation for FFPE Visium libraries. Sequencing cycles comprised 28 bp (Read 1), 10 bp (Index 1 and 2) and 50 bp (Read 2). Illumina NextSeq 2000 sequencing was performed using NextSeq 1000/2000 P2 reagents (100 cycles), whereas GeneMind Genolab M used FCM reagents (150 cycles).

Bioinformatics processing

Raw sequencing reads were primary processed in the Space Ranger v.1.3 using the count command with default parameters. Raw reads were additionally assessed with FastQC [15] algorithm, and results were visualized with MultiQC [16] tool. After read mapping and counting, resulted feature-barcode matrix was uploaded into R environment and processed via the Seurat [17] R package. The Pearson correlation coefficient was used to evaluate consistence in total UMI and detected gene number between two platforms in tissue spots. Overlapping of the genes between two platforms was visualized using Venn diagram. Normalization of the raw read counts in feature-barcode matrix was performed with the SCTransform [18] function in the Seurat package. The Pearson correlation coefficient between normalized expression profiles of overlapped genes was calculated for each tissue spot in the samples. The SCTransform normalized counts were merged and renormalized without application of batch effect correction. Linear dimension reduction via principal component analysis (PCA) and non-linear dimension reduction via uniform manifold approximation and projection [19] (UMAP) were applied to the merged normalized counts. Clusterization of tissue spots was performed with shared nearest neighbor (SNN) clusterization using first 30 principal components. FindAllMarkers function in the Seurat package was applied to SCT normalized counts using the Wilcoxon Rank Sum test to reveal differentially expressed genes (DEGs) in spatial clusters for each sample separately. FDR correction was performed using the Benjamini–Hochberg method. The distributions of log twofold-change (LFC) and false discovery rate (FDR) of DEGs were used to assess non-overlapping DEGs. The ggplot2 [20] R package was used to visualize most of comparison results.

Availability of data and materials

The data underlying this article are available in the article and in its online supplementary material. Raw data are available in NCBI BioProject repository via accession number PRJNA894440. The script was submitted to Github and can be found through the following link: https://github.com/piams/Comparison-of-the-NextSeq-2000-and-Genolab-M-for-spatial-transcriptomics.

References

  1. Karlsson M, Zhang C, Méar L, Zhong W, Digre A, Katona B, et al. A single–cell type transcriptomics map of human tissues. Sci Adv. 2021;7:eabh2169.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  2. Regev A, Teichmann SA, Lander ES, Amit I, Benoist C, Birney E, et al. The human cell atlas. Elife. 2017;6:e27041.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Luecken MD, Theis FJ. Current best practices in single-cell RNA-seq analysis: a tutorial. Mol Syst Biol. 2019;15:e8746.

    Article  PubMed  PubMed Central  Google Scholar 

  4. Tang X, Huang Y, Lei J, Luo H, Zhu X. The single-cell sequencing: New developments and medical applications. Cell Biosci. 2019;9:1–9.

    Article  Google Scholar 

  5. Adil A, Kumar V, Jan AT, Asger M. Single-cell transcriptomics: current methods and challenges in data acquisition and analysis. Front Neurosci. 2021;15:398.

    Article  Google Scholar 

  6. Williams CG, Lee HJ, Asatsuma T, Vento-Tormo R, Haque A. An introduction to spatial transcriptomics for biomedical research. Genome Med. 2022;14:1–18.

    Article  Google Scholar 

  7. Ståhl PL, Salmén F, Vickovic S, Lundmark A, Navarro JF, Magnusson J, et al. Visualization and analysis of gene expression in tissue sections by spatial transcriptomics. Science (80-). 2016;353:78–82.

    Article  Google Scholar 

  8. Moses L, Pachter L. Museum of spatial transcriptomics. Nat Methods. 2022;19:534–46.

    Article  CAS  PubMed  Google Scholar 

  9. Levy SE, Myers RM. Advancements in next-generation sequencing. Annu Rev Genomics Hum Genet. 2016;17:95. https://doi.org/10.1146/annurev-genom-083115-022413.

    Article  CAS  PubMed  Google Scholar 

  10. Anslan S, Mikryukov V, Armolaitis K, Ankuda J, Lazdina D, Makovskis K, et al. Highly comparable metabarcoding results from MGI-Tech and Illumina sequencing platforms. PeerJ. 2021;9:e12254.

    Article  PubMed  PubMed Central  Google Scholar 

  11. Li C, Fan X, Guo X, Liu Y, Wang M, Zhao XC, et al. Accuracy benchmark of the GeneMind GenoLab M sequencing platform for WGS and WES analysis. BMC Genomics. 2022;23:533.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  12. Liu Y, Han R, Zhou L, Luo M, Zeng L, Zhao X, et al. Comparative performance of the GenoLab M and NovaSeq 6000 sequencing platforms for transcriptome and LncRNA analysis. BMC Genomics. 2021;22:829.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  13. Andersson A, Larsson L, Stenbeck L, Salmén F, Ehinger A, Wu SZ, et al. Spatial deconvolution of HER2-positive breast cancer delineates tumor-associated cell type interactions. Nat Commun. 2021;12:1–14.

    Article  Google Scholar 

  14. Chen F, Dong M, Ge M, Zhu L, Ren L, Liu G, et al. The history and advances of reversible terminators used in new generations of sequencing technology. Genomics Proteomics Bioinformatics. 2013;11:34–40.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Andrews S. FastQC: a quality control tool for high throughput sequence data. 2010. Available online at: https://www.bioinformatics.babraham.ac.uk/projects/fastqc. Accessed 11 Jul 2021.

  16. Ewels P, Magnusson M, Lundin S, Käller M. MultiQC: summarize analysis results for multiple tools and samples in a single report. Bioinformatics. 2016;32:3047–8.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  17. Hao Y, Hao S, Andersen-Nissen E, Mauck WM, Zheng S, Butler A, et al. Integrated analysis of multimodal single-cell data. Cell. 2021;184:3573-3587.e29.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  18. Hafemeister C, Satija R. Normalization and variance stabilization of single-cell RNA-seq data using regularized negative binomial regression. Genome Biol. 2019;20:1–15.

    Article  Google Scholar 

  19. Becht E, McInnes L, Healy J, Dutertre CA, Kwok IWH, Ng LG, et al. Dimensionality reduction for visualizing single-cell data using UMAP. Nat Biotechnol. 2018;37:38–44.

    Article  Google Scholar 

  20. Wickham H. ggplot2. ggplot2. Elegant graphics for data analysis. NY: Springer New York: 2009. https://doi.org/10.1007/978-0-387-98141-3.

Download references

Acknowledgements

Visium library preparation and sequencing were carried out on the equipment of the Core Facility "Medical Genomics" (Tomsk NRMC) and Genomed Ltd.

Funding

This work was supported by the Russian Science Foundation under research project №21–75-10021, by Tomsk State University Development Programme (Priority-2030) and by the RUDN University Strategic Academic Leadership Program.

Author information

Authors and Affiliations

Authors

Contributions

PI performed bioinformatics analysis, analyzed and interpreted the data. IL performed the histological analysis, IL and TG performed visium library preparation. PI and IL were major contributors in writing the manuscript. DP, PhK and SK performed sequencing using GeneMind Genolab M sequencing platform. ED performed sequencing using Illumina NextSeq 2000 sequencing platform and initiated the study. All authors read and approved the final manuscript. Authors agreed to be personally accountable for the author's own contributions.

Corresponding authors

Correspondence to Larionova Irina or Denisov Evgeny.

Ethics declarations

Ethics approval and consent to participate

The study was carried out according to Declaration of Helsinki and was approved by the local committee of Medical Ethics of Tomsk Cancer Research Institute; all patients signed informed consent for the study.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1: Supplementary Table S1. 

FastQC report combined and retrieved via MultiQC.

Additional file 2: Figure S1.

A. Overlap of detected genes. B, C. Raw read count distribution of the common and unique genes. Count is a number of genes. Figure S2. Raw read count distribution of the common and unique DEGs in the clusters. Y-axis is a number of DEGs. Figure S3. Box plots describing difference in –log (FDR) and LFC between overlapping (Common) and platform-unique (Unique) DEGs.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Pavel, I., Irina, L., Tatiana, G. et al. Comparison of the Illumina NextSeq 2000 and GeneMind Genolab M sequencing platforms for spatial transcriptomics. BMC Genomics 24, 102 (2023). https://doi.org/10.1186/s12864-023-09192-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12864-023-09192-w

Keywords